Description

In the default configuration, the archive accepts DICOM associations from an unregistered calling AE. A remote peer can use C-FIND and C-GET to enumerate and retrieve patient information and medical images. It can also use C-STORE to file fabricated study content under an existing patient identity.

Traduire en français

Produits concernés

Scores CVSS

VersionScoreSourceVecteur
4.0 9.3 Critique ics-cert@hq.dhs.gov CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X
3.1 9.1 Critique ics-cert@hq.dhs.gov CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N