CVE-2019-25719

8.6 Élevée

Publiée le 02/06/2026 à 16:16 · Modifiée le 22/07/2026 à 21:10 · En attente d'analyse NVD · Source : disclosure@vulncheck.com

Description

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.