Veille CVE

Recherche « Draeger delta xl firmware »

2 CVE

  1. CVE-2019-25717 4.3 Moyenne

    Dräger Infinity Delta · Dräger Infinity Delta XL · Dräger Infinity Kappa · Draeger infinity delta firmware

    Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain an information disclosure vulnerability that allows unauthenticated network attackers to access log files over a network connection. Attackers can retrieve device internals, location information, and wired network configuration details from the exposed log files.

  2. CVE-2019-25716 7.5 Élevée

    Dräger Infinity Delta · Dräger Infinity Delta XL · Dräger Infinity Kappa · Draeger infinity delta firmware

    Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that allows remote attackers to cause the monitor to reboot by sending a malformed network packet. Attackers can repeatedly send malformed network packets to disrupt patient monitoring until the device falls back to default configuration and loses network connectivity.